Every Pro Display XDR and Apple Studio Display undergoes a state-of-the-art factory display calibration process on the assembly line to ensure the accuracy of the P3 wide color panel and LEDs.
The factory display calibration process lets Pro Display XDR and Studio Display users enjoy an exceptional viewing experience out of the box. If your workflow requires custom calibration, you can fine-tune the calibration of your display, or recalibrate it for a workflow tuned to a third-party spectroradiometer.
Measure your display
You can use a set of QuickTime movie test patterns from Apple to evaluate the calibration of your Pro Display XDR or Studio Display. These appropriately color-tagged references allow you to use your in-house spectroradiometer to measure and verify the color primaries/secondaries and luminance including the electro-optical transfer function (EOTF).
Check compatibility
- For Pro Display XDR, make sure that your Mac is using macOS 10.15.6 or later and your display uses Display Firmware 4.2.37.
- For Studio Display, make sure that your Mac is using macOS 13.3 or later and your display uses Studio Display Firmware Update 16.4.
Download test patterns
- Go to the AVFoundation Developer Page.
- In the Related Resources section, click Color Test Patterns to download the test files.
- Double-click the QuickTime-Test-Pattern.zip file to decompress the archive.
Measure the display
- Align the third-party spectroradiometer to the center of the display.
- Make sure the room is dim or dark to prevent stray light or glare from affecting the measurements.
- Open the QuickTime Test Pattern Movies folder and choose the set of patterns that you want to test. Each folder contains sequences of movie files for measuring color or luminance in configurations such as BT.709 and BT.601.
- Open each file in QuickTime Player and measure each test pattern movie file in the folder. Make sure that the reference preset currently in use matches the chosen test pattern. For example, use the Digital Cinema (P3–DCI) mode when using the DCI SDR-based patterns.
- Compare the color (chromaticity) and luminance values you measured to those in the Reference Values.txt file in the test pattern’s folder. Depending on the tolerance or calibration of your spectroradiometer, there may be some variation in readings relative to the reference values.
If your measurements vary significantly from the reference values, get tips for measuring your display.
Fine-tune the calibration
If your workflow is tuned to a specific target, you can fine-tune the calibration of your Pro Display XDR or Studio Display. This lets you adjust the white point and luminance of your display to more precisely match your own in-house display calibration target.
Perform in-field recalibration
If you need to recalibrate Pro Display XDR or Studio Display for a workflow tuned to a third-party spectroradiometer, you can use the Display Calibrator app included with macOS Ventura 13.3 or later. If you have an earlier version of macOS, you can download the Pro Display XDR Calibrator app to recalibrate Pro Display XDR on a Mac with macOS Catalina 10.15.6 or later.
Tips for measuring and calibrating your display
If you aren't getting consistent results measuring and calibrating your display, learn what to do.
Check your environment
- Control the ambient environment to minimize stray light reflecting onto the aperture target area on the display.
- Apple recommends measuring and calibrating in an environment with ambient temperatures of 77 degrees fahrenheit or cooler. For the most repeatable results, ambient temperatures should be similar during calibration and in typical use.
Check the setup of your spectroradiometer
- Make sure the display has warmed up for at least 30 minutes and is thermally stable.
- Orient the meter so that it's perpendicular with the display and level on the x-axis to minimize polarization.
- Check your meter’s exposure setting. If your meter has an auto-exposure mode, turn it on—this optimizes readings at various luminance levels. If your meter doesn’t have an auto exposure mode, check with the instrument's manufacturer for recommended settings. Review configuration options for positing and focal distance, aperture, bandwidth, exposure, measurement speed, display sync, and any automatic or dynamic features your instrument may have.
- Check if your meter has an integration time setting. Use a longer integration time when possible for a more accurate result.
- If your meter has an auto neutral-density (ND) filter setting, turn it off.
Check the calibration of your spectroradiometer
Check the manufacturer’s specifications to determine your spectroradiometer’s tolerance. Observed variance may result from any of the following conditions:
- Different instrument calibrations that use different incidences of illuminant A.
- Uncertainty inherent in each individual illuminant A.
- Different instrument hardware design and optical pathways, resulting in different inherited errors.
Reset the spectroradiometer
If your instrument becomes unresponsive, contact the manufacturer for instructions on how to reset the spectroradiometer.
Learn more
Learn about the macOS reference presets included with your display.